3. To set another electrode rotating speed at the end of the method, select “D/AOUT final” and enter, in mV, a DC constant potential between 0 and +5000 mV tobe output at the PGZ or PST050 “D/A OUT” socket at the end of the method. Thesame principle as for the “D/A OUT initial” is applied. Note that the rotation speedis set to zero by default at the end of the method (“D/A OUT final = 0”).4. Run the experiment (command “Run” - “Start”, or icon or F5 function key).The EDI101 or EDI10000 starts rotating at the selected speed when the methodrun is initiated then stops rotating at the end of this method.152VoltaLab 21/06/10/40/50/80Section 6, Chapter 1: Rotating Disc Electrode (EDI)