3.7 Set up the Other settings parametersIf you selected the Routine setup mode while installing VoltaMaster 4, the Othersettings are entered at the end of the software installation and cannot be changedIf you want to change these parameters afterwards, you must switch to anotherVoltaMaster 4 setup mode. You do not have to re-install the software to change thesetup mode.Press F1 to consult the Help file, keyword index:How to select/change the VoltaMaster 4 setup modes?.If you selected the Full or Supervisor/Operator setup modes while installingVoltaMaster 4, select the "Settings" menu then the "Other settings" command.. If you are using a VoltaLab 21 (PGP201):Check that "Display curve at end - Raw data" and "Display curve at end -Curve data" are selected.The Other settings are not significant when using a VoltaLab 21.. If you are using a VoltaLab 06, 10, 40, 50 or 80:Check that the "Anti-oscillation filter" is cleared. Select the frequency (50 Hz or60 Hz) of the power supply. This selection is used by VoltaMaster 4 to minimizethe background noise ratio on the measurements. Check that the otherparameters are as follows:VoltaLab 21/06/10/40/50/80 35Section 3, Chapter 7: Set up Other settings parameters