If it is not the case (communication failure (1) message), click the “Cancel” button,check the Instrument setup parameters and connections (instruments powered,High Current Booster Cell button depressed), click the “Settings” then the “Test”buttons.Important:VoltaMaster 4 saves the High Current Booster settings when you close theInstrument setup box with the “OK” button. Then, each time you run a sequence,VoltaMaster 4 will search for the HCBXXX connection then configure the HCBXXXelectrode connection as stated in the HCBXXX settings. Thus, proceed as followswhen you want to run a sequence:. enter/check the HCBXXX settings,. connect the electrodes,. click “OK” to close the High Current Booster Settings box then “OK” again toclose the ”Instrument setup” box. VoltaMaster 4 tests the connection to thePGZ402 and HCB and configurates the HCBXXX electrode connection asstated in the “High Current Booster Settings” box.. before running the sequence, switch on the “Cell” button on the HCBXXX.Do not switch on the “Cell” button on the PGZ402!. run the sequence.For more information, see section 3 “Getting started” or consult the Help file,keyword indexes:Instrument setup and High Current Booster settings.184VoltaLab 21/06/10/40/50/80Section 6, Chapter 8: HCB005 - HCB010 - HCB020 High Current Boosters