14. Sequence emptyYou are trying to run an experiment having no method programmed inside. Checkthe sequence of methods (command “Sequence - Sequence edition”).15. The detected instrument does not correspond to the one selectedThis error message is displayed upon starting a sequence if you have selectedan instrument type that does not correspond to the one connected to the PC.Select “Settings – Instrument setup” or “Edit - Instrument setup” then check the“Potentiostat” entry. The option selected must correspond to the potentiostat youhave connected to <strong>your</strong> PC.16. Thermic disjunctionThis error message indicates that an overheat of the power circuits of the PGZ orPST potentiostat has occurred. The sequence in progress is stopped.· The room temperature must be between 5 and 40°C. Move the potentiostataway from any source of heat. The PGZ or PST potentiostat must be placedin a properly ventilated place: leave enough place behind the potentiostat rearpanel.· Wait for the CELL button lamp to light up (the CELL lamp lights up as soonas the overheat of the power circuit is away) and restart the sequencefrom VoltaMaster 4. If the problem remains, contact <strong>your</strong> HACH LANGErepresentative.VoltaMaster 4 - Version 7.0x 95Section 4, Chapter 8: Error messages