Declare the AMU130 as a “Potentiostat (X/Y/Vg)” in the “Additional unit”.Click the “Test” button. The instrument type (PGZ402) and version are displayed.If it is not the case (communication failure (1) message), select “VoltaLab PGZ402”for “Potentiostat”, select the communication port (1 to 8) of the PC used for theconnection to the PGZ402 for “Serial port” then click again the “Test” button.On the AMU130 perform the following settings:. Power switch to “On”.. Potential switch to “Ext.”.. “Cell Ext./Int.” switch to “Int.” to use the AMU130 built in dummy cell or “Ext.” touse a real cell with 2 electrodes (REF and WORK) connected to the AMU130.. Range switch to “200 pA” as an example. This switch sets the AMU130 currentmeasurement range.Use with a real cell (electrodes)POTENTIALCELLEXT. +0-1V WORK WEEXT.POWERONOFFOFFSETCURRENTFILTERAMU130INT.RANGE2nA±1V 20nA1ms 10200pAREF.electrodeREFelectrodeUse with the AMU130 dummy cellPOTENTIALEXT. +0-1VPOWERCURRENTON OFFSETFILTERAMU130CELLWORKEXT.INT.RANGE REF.2nA200pAOFF±1V 20nA1ms 10VoltaLab 21/06/10/40/50/80 193Section 6, Chapter 10: AMU130 - PGZ402 E(X) IN and I(Y) IN sockets