5.2.3 InstallationElectrical safety• Do not connect the instrument to any electrical source that uses a 230V ITneutral regime• A bipolar circuit breaker must be installed in a 2-phase mains power supplywithout neutral• Always disconnect the instrument before any intervention• The instrument must be connected to an electrical system which complies withapplicable local regulations• All cables connected to the instrument must be fire resistant, type UL94V-1• This instrument is intended for indoor use only. Under no circumstances shouldit be used outdoorsWarning!In case of flammable liquid being sprayed on the instrument, use the local circuitbreaker to turn off the instrument (not the on/off switch), clean immediately and getsome fresh air.Warning!Set up the instrument in a properly ventilated place. The power supply connectorson the front panel must remain accessible so that you can quickly disconnect thecables in case of emergency. The room temperature should not exceed 40°C.Warning!Electrical dangers and fire hazard.Only use the supplied power cable.Only qualified experts may perform the tasks described in this section of themanual, while adhering to all locally valid safety regulations.Warning!Removeable power cables must not be replaced with inadequately dimensionedpower cables.140VoltaLab 21/06/10/40/50/80Section 5, Chapter 2: The PGP201 Potentiostat/Galvanostat

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!