4.3.46 Standard calibrationThis method calculates the concentration taking in account dilution volumesand handling up to 4 calibration curves (independent setting for their respectiveconcentrations) plus a blank subtraction. The signals can be a peak integration(Total, positive or negative I or dI) or a peak coordinate in potential, differentialpotential, charge, current, differential current or A/D IN with selectable potential,time or current windows. Standard calibration is a common protocol to determinemedium concentrations. The principle is to run a set of standard solutions ofwell k<strong>now</strong>n concentration and then to run the sample of unk<strong>now</strong>n concentration.Interpolation provides the concentration in the sample, assuming that the recordedsignal is proportional to the concentration and handling dilution volumes if needed.Linear, as well as logarithmic or exponential relationships are handled.Consult the Help file, keyword index: Standard calibration - Method editing.VoltaMaster 4 - Version 7.0x 61Section 4, Chapter 3: Methods available