02.08.2013 Views

FONIX® FP35 Touch - Frye Electronics

FONIX® FP35 Touch - Frye Electronics

FONIX® FP35 Touch - Frye Electronics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Chapter 4 97<br />

Automated Test Sequences<br />

There are several different automated test sequences available with the <strong>FP35</strong><br />

analyzer. One test sequence of your choice is included as a standard feature<br />

when the <strong>FP35</strong> is ordered. The other test sequences can be added as optional<br />

features.<br />

The available automated test sequence include: ANSI 96, ANSI 03, IEC 94, IEC<br />

05, JIS, and ISI.<br />

4.1 The ANSI Sequence<br />

The ANSI test sequence allows you to test hearing aids according to the ANSI<br />

S3.22 standard. You can use the ANSI test to control the quality of the hearing<br />

aids that you dispense. Compare the manufacturer’s specifications with your<br />

own ANSI measurements of an aid. If they do not conform within expected tolerances,<br />

you can contact the manufacturer.<br />

There are two versions of the ANSI S3.22 standard on the <strong>FP35</strong> analyzer: ANSI<br />

96 and ANSI 03. Check your hearing aid specification sheet to determine which<br />

standard was used on your hearing aid.<br />

Here are the major differences in the two versions of the standard:<br />

• When you set up an AGC aid for an ANSI 03 test sequence, you will start<br />

with its compression controls set to minimum (with the compression<br />

knee point set as high as possible). Just before the input/output measurements,<br />

the test sequence will pause to allow you to set the compression<br />

controls of the aid to maximum (with the compression knee point set as<br />

low as possible). In ANSI 96, AGC aids are tested with their compression<br />

controls set to maximum for all measurements.<br />

• When you adjust the gain control of the hearing aid midway through the<br />

ANSI 03 automated test sequence, the measured reference test gain value<br />

needs to be within 1.5 dB of the target value. In ANSI 96, the measured<br />

value needs to be within 1 dB of the target value.<br />

• The EIN formula in ANSI 03 uses a 50 dB SPL input instead of the 60 dB<br />

SPL input used in ANSI 96. With ANSI 96, any aid with a compression<br />

knee point below 60 dB SPL showed artificially high EIN results. This<br />

means that ANSI 03 EIN test results should be better (lower) for AGC<br />

aids.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!