22 FONIX <strong>FP35</strong> Hearing Aid Analyzer<br />

You can always tell what type a variable is by looking at the dash sequence next<br />

to it in one of the menus. Local variables use a series of small dots. Global variables<br />

use a series of large dots. Resetable variables use a series of short dashes.<br />

See Figure 2.1.2.<br />

Figure 2.1.2—Local menu of the Coupler Multicurve screen<br />

Resetable<br />

variable<br />

Local<br />

variable<br />

Global<br />

variable<br />

Details on the behavior of menu items:<br />

Local variables are not available in the Default Settings menu. When the analyzer<br />

is first powered on, local variables are set to a factory defined default setting,<br />

but they are not normally changed back to the default setting when [RESET] is<br />

pressed. (There are some exceptions.)<br />

When you change a resetable variable in the Default Settings menu, it will<br />

impact that variable in all local menus where it is available. However, changing<br />

a resetable variable in a local menu will not change its value in the Default<br />

Settings menu. When [RESET] is pressed, the local menu variable will revert<br />

back to the value defined in the Default Settings menu.<br />

Changing a global variable will change its value in all screens where it is available,<br />

including the Default Settings menu.<br />

2.1.3 Using the Help Menus<br />

Each screen of the <strong>FP35</strong> analyzer contains a pop-up help feature to aid you if<br />

you ever get stuck. If you find yourself uncertain of your next step, press the