02.08.2013 Views

FONIX® FP35 Touch - Frye Electronics

FONIX® FP35 Touch - Frye Electronics

FONIX® FP35 Touch - Frye Electronics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

160 FONIX <strong>FP35</strong> Hearing Aid Analyzer<br />

3. Press [NEXT] twice to enter the Custom menu.<br />

4. Make sure that DISTORTION is OFF. Use the arrow keys to make this selection<br />

if necessary.<br />

5. Use [,] to select IM FREQ DIFF.<br />

6. Use [,] to select the fixed distance between the primary and secondary frequencies<br />

used in the measurement. You can choose from no distance (HARM)<br />

to 4000 Hz.<br />

7. Press [EXIT] to return to the real-ear measurement screen.<br />

8. Press and hold down [F4]. This will pop up a list of the available selections,<br />

including DIST.<br />

9. Use [,] to select DIST.<br />

10. Press [START/STOP] to make the selection.<br />

11. Use [,] to select the RMS source of the test. Each frequency will have an<br />

amplitude 3 dB less than the RMS source.<br />

12. Set up the client, the aid, and the analyzer for real-ear testing. Make sure to<br />

level the sound field speaker.<br />

13. Press [START/STOP] to start the static IM distortion measurement. See Figure<br />

5.7.3.1.<br />

14. Use [,] to change the primary frequency of the test. The secondary frequency<br />

will also change, remaining a fixed distance (selected in Step 6) away from<br />

the primary frequency.<br />

15. Press [START/STOP] to stop the measurement.<br />

Figure 5.7.3.1—Real-Ear static IM distortion test

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!