02.08.2013 Views

FONIX® FP35 Touch - Frye Electronics

FONIX® FP35 Touch - Frye Electronics

FONIX® FP35 Touch - Frye Electronics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

148 FONIX <strong>FP35</strong> Hearing Aid Analyzer<br />

Figure 5.5.1—SPL display<br />

5.5.2 Running an SPL Test<br />

This section assumes you have already entered the client’s audiometric data in<br />

the Audiogram Entry Screen, and that you have generated a target. See Section<br />

5.3.3 and 5.3.4 for more details.<br />

1. Place the probe tube into the client’s ear following the instructions provided<br />

in Section 5.2.4.<br />

2. Insert the aid into the client’s ear, being careful not to drag the probe tube<br />

farther into the ear canal (See Figure 5.2.4C). Do not turn the aid on.<br />

3. Instruct the client to remain still when the signal is on.<br />

4. Press [F5] to level.<br />

5. Have the client turn the aid on and set the gain control to the normal use<br />

level.<br />

6. Use [F2] to select AIDED 1.<br />

7. Use [F4] to select the source type for AIDED 1, and use the [,] keys to set<br />

the source amplitude<br />

8. If necessary, adjust noise reduction and output limiting with [MENU].<br />

9. Have the client resume the position they were in during leveling, and<br />

remind them to stay still while the signal is on.<br />

10. Press [START/STOP] to test. If you are using the COMP or DIG SPCH signal<br />

source, you will need to press [START/STOP] again to stop the signal.