178 FONIX <strong>FP35</strong> Hearing Aid Analyzer<br />

The sections in this chapter will explain the following:<br />

• 6.1: Target creation<br />

• 6.2: Displays<br />

• 6.3: Measurements<br />

• 6.4: Reference<br />

6.1 Target Creation<br />

In order to create a target for use in the Coupler Target screen or the Coupler<br />

EarSim screen, you must enter the Audiogram Entry screen, just as you would if<br />

you were creating a normal real-ear target.<br />

1. Enter the Audiogram Entry screen by pressing [F2] from the Opening screen<br />

and using the [NEXT] and [BACK] keys if necessary.<br />

2. Follow the instructions found in Section 5.3 to enter the patient’s audiogram<br />

and to create a real-ear target.<br />

3. Follow the instructions found in Section 5.3.6, if desired, to measure the<br />

patient’s RECD.<br />

4. Press [EXIT] to return to the Opening screen.<br />

5. Press [F3] to enter one of the coupler multicurve screens. Use the [NEXT]<br />

and [BACK] keys to enter the Coupler Target screen or the Coupler EarSim<br />

screen.<br />

6. Press [MENU] and [BACK] to enter the Target menu. Set the FIT RULE, AGE,<br />

AID TYPE, VENT, and TUBING (if applicable) in this menu using the arrow<br />

keys. If you select NAL-NL1 or MOD NAL as your FIT RULE, you will also<br />

have to set COMPRESSION, CHANNELS, AID LIMITING, and FIT TYPE.<br />

See Section 5.3.4 for details.<br />

7. Press [EXIT] to return to the measurement screen. You are now ready to perform<br />

the measurements and compare them to the displayed target.<br />

The Coupler Target screen converts the real-ear target into a coupler target displayed<br />

in either dB GAIN or dB SPL. The Coupler EarSim screen displays realear<br />

targets in either dB IG (insertion gain) or dB SPL. When the Coupler EarSim<br />

display is set to dB SPL, the patient’s HTL and UCL values are also displayed.<br />

To switch between the gain and SPL displays in each screen, use the [MENU]<br />

key to open the local menu, and use the arrow keys to select GAIN or SPL in the<br />

DISPLAY field. [EXIT] will close the local menu.<br />

Targets are usually tied to the source type and source level of CRV 2. If you<br />

would like to tie the target to a different curve, or to the currently selected<br />

curve, change the TARGET REF setting in the Target menu. See Section 6.3.3.<br />

If you have modified the actual numerical target values in the Audiogram Entry<br />

screen, or entered your own target entirely, you will not be able to enter the