Real-Ear Measurements 141<br />

4. Use the [,] keys to adjust the source amplitude to the desired level, usually<br />

the same level used for the unaided measurement.<br />

5. Instruct the client to resume the position they were in for leveling and testing.<br />

Tell them to sit very still while the signal is on.<br />

6. Press [START/STOP] to introduce the signal. The aided response will appear<br />

on the graph as a thin line labeled with a curve number.<br />

Note: If you are using the COMP or DIG SPCH signal, you can change the<br />

source amplitude during testing to see how the aid responds to such changes.<br />

7. Press [START/STOP] to end the measurement if you are using the COMP or<br />

DIG SPCH signal.<br />

8. If desired, press [F2] to select a another curve number and repeat steps 3-7.<br />

See Figure 5.4.2A to view the aided response curve in terms of aided gain. See<br />

Figure 5.4.2B to view the aided response in terms of insertion gain.<br />

Figure 5.4.2A—Aided Gain screen. The N is the unaided curve. The 1 and 2 curves represent<br />

aided curves 1 and 2. The stars indicate the aided gain target.