Fitting Formula Tables 225<br />

1/3, 1/2, 2/3 Gain<br />

The 1/3-2/3 Insertion Gain Hearing Aid Selection Guide<br />

by Robert Libby<br />

Hearing Instruments vol 37 #3, 1986<br />

Required insertion gain is calculated as follows:<br />

FREQ. INS GAIN INS GAIN INS GAIN<br />

(Hz) (dB) (dB) (dB)<br />

(2/3 rule) (1/3 rule) (1/2 rule)<br />

250 2/3 HTL - 5 dB 1/3 HTL - 5 dB 1/2 HTL -5 dB<br />

500 2/3 HTL - 3 dB 1/3 HTL - 3 dB 1/2 HTL -3 dB<br />

750* 2/3 HTL 1/3 HTL 1/2 HTL<br />

1000 2/3 HTL 1/3 HTL 1/2 HTL<br />

1500* 2/3 HTL 1/3 HTL 1/2 HTL<br />

2000 2/3 HTL 1/3 HTL 1/2 HTL<br />

3000 2/3 HTL 1/3 HTL 1/2 HTL<br />

4000 2/3 HTL 1/3 HTL 1/2 HTL<br />

6000 2/3 HTL - 5 dB 1/3 HTL - 5 dB 1/2 HTL -5 dB<br />

8000* 2/3 HTL - 5 dB 1/3 HTL - 5 dB 1/2 HTL -5 dB<br />

Note: Frequencies marked by an asterisk ( * ) are interpolated because the article<br />

does not reference them.