Real-Ear Measurements 149<br />

11. Repeat steps 6 through 10 for AIDED 2 and AIDED 3.<br />

In general, AIDED 1 is tested at low levels (50 dB SPL) to make sure soft sounds<br />

are above the patient’s threshold values. AIDED 2 is tested at medium levels (65-<br />

70 dB SPL) to make sure that the aid meets the target. AIDED 3 is tested at loud<br />

levels (90 dB SPL) to make sure that loud sounds are beneath the patient’s UCLs.<br />

Note: See Section 5.4.5 for information on testing open fit hearing aids.<br />

Figure 5.5.2—SPL display with measurements<br />

5.5.3 Viewing Real-ear SPL Results in Full Screen<br />

Normally the Real-ear SPL screen is displayed with the graph on the left important<br />

information on the right. You can remove the information and view the<br />

results on a large wide graph as shown in Figure 5.5.3.<br />

To do this:<br />

1. Press [MENU].<br />

2. Use the arrow keys to change the DATA/GRAPH to GRAPH 2X.<br />

3. Press [MENU] to close the menu. The graph will now take up the entire display.<br />

Alternately, you can use the left and right arrow keys in the Real-ear SPL screen<br />

to quickly switch between the standard Real-ear SPL display, the full-sized Realear<br />

SPL display, and the numerical data display.