110 FONIX <strong>FP35</strong> Hearing Aid Analyzer<br />

10. OSPL90 REF: frequency response of OSPL90 curve at reference frequency<br />

(1600 or 2500 Hz). Select with [F3].<br />

11. OSPL90 MAX: amplitude and frequency of the maximum output of the aid.<br />

12. FOG CALC: target reference test gain.<br />

13. FOG MEAS: measured reference test gain.<br />

14. EIN: equivalent input noise measurement.<br />

15. LIM: response limit level determined by taking the frequency response of the<br />

RESP 50 (or 60) curve at the reference frequency and subtracting 15 dB.<br />

F1: frequency at which the RESP 50 (or 60) curve crosses the response limit<br />

level on the way up.<br />

F2: frequency at which the RESP 50 (or 60) curve crosses the response limit<br />

level on the way down.<br />

16. TOT H DIST: total harmonic distortion measurements and the frequencies<br />

and amplitudes at which they are taken.<br />

17. Battery size and current measurement.<br />

18. I/O CURVE: input/output curve at reference frequency (1600 or 2500 Hz).<br />

Select with [F3].<br />

Figure 4.3.1A—JIS test results, Frequency Response Screen