02.08.2013 Views

FONIX® FP35 Touch - Frye Electronics

FONIX® FP35 Touch - Frye Electronics

FONIX® FP35 Touch - Frye Electronics

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Chapter 7 191<br />

The Audiometer Option<br />

7.1 Introduction<br />

The audiometer option on the <strong>FP35</strong> turns your hearing aid analyzer into a type<br />

5 screening audiometer. In other words, you can use the <strong>FP35</strong> for doing puretone<br />

air conduction screening using either insert earphones or sound field testing.<br />

It is not intended for use as a clinical audiometer.<br />

7.2 Equipment<br />

The audiometric testing is performed using either a sound field speaker or one<br />

or two insert earphones. If you use two earphones, you can set your client up<br />

with two earphones and perform your measurements one ear at a time. If you<br />

use one earphone, take the set of measurements for one ear, then switch the<br />

earphone into the other ear and take the other set of measurements. TDH headphones<br />

are not supported on the <strong>FP35</strong>.<br />

7.3 Setup<br />

The following describes how to set up the <strong>FP35</strong> for audiometric measurements.<br />

If you are using the sound field speaker as the transducer, you must first level.<br />

This is described in Section 7.3.1. The insert earphone setup is described in<br />

Section 7.3.2.<br />

7.3.1 Sound Field Speaker<br />

There are two methods to leveling the sound field speaker for audiometric measurements.<br />

The first method, called the head surface method, is identical to the<br />

leveling method used for real-ear measurements – it is performed with the client<br />

present and the reference microphone in position above the client’s ear. The<br />

second method, called the free field method, is performed with the microphone<br />

positioned in front of the sound field speaker without the presence of the client.<br />

Head Surface Method<br />

1. Set up your <strong>FP35</strong> analyzer as you would for real-ear measurements with the<br />

sound field speaker in place. See Section 9.2.<br />

2. Situate your client with the sound field speaker at a 45º or 0º angle azimuth.<br />

See Figure 7.3.1A. We recommend a testing distance of 12 inches from the<br />

sound field speaker. You can test at longer distances but it will limit the<br />

maximum output possible.<br />

3. Enter the Audiometric Entry Screen by pressing [F2] and using the [NEXT]<br />

and [BACK] keys as necessary.<br />

4. Press [MENU].<br />

5. Select EAR and choose the desired ear.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!