Real-Ear Measurements 135<br />
5.3.9 Starting a New Client<br />
When you have finished performing the real-ear (or target coupler) measurements<br />
on the hearing aid and patient, pressing [F1] in the Audiogram Entry<br />
screen will erase all measurements and restore settings in order to make setting<br />
up for a new patient quick and easy.<br />
In addition to all audiometric and real-ear measurements being erased, the<br />
following menu items are reset when [F1] is pressed in the Audiogram Entry<br />
screen. They are also set when the analyzer is powered on or when the [RESET]<br />
key is pressed.<br />
• CHANNELS – 1 (used by NAL-NL1 and MOD NAL)<br />
• AID LIMITING – Wideband (used by NAL-NL1 and MOD NAL)<br />
• FIT TYPE – Bilateral (used by NAL-NL1 and MOD NAL)<br />
• VENT – Tight (used by coupler targets only)<br />
• TUBING – Libby #4 (used by coupler targets only)<br />
• SOUND FIELD – 45 degrees<br />
• TARGET REF – REF #2<br />
• TABULATION – Target<br />
• REF METHOD – HEAD SURFACE<br />
The following menu items are set to their default values only when the analyzer<br />
is first powered on. They aren’t reset when NEW CLIENT is pressed or when<br />
[RESET] is pressed. Their default values can be changed in the Default Settings<br />
menu. See Section 2.3.1. (The listed values are the factory defaults.)<br />
• FIT RULE – NAL-NL1<br />
• AGE – Adult<br />
• ASSESSMENT – Ins. Ear<br />
• COMPRESSION – 52 dB (used by NAL-NL1, DSL, and MOD NAL only)<br />
• AID TYPE – ITC<br />
• AID GROUP – Standard<br />
• REF MIC – ON<br />
• COUPLER TYPE – 2 cc (used by Coupler measurement screens only)