(8.5x11) 6000 Manual Cover - Newport Corporation
(8.5x11) 6000 Manual Cover - Newport Corporation
(8.5x11) 6000 Manual Cover - Newport Corporation
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Electrical Specifications:<br />
DC Current Measurement (Low-Power, Semiconductor Photodiode CAL MODULE)<br />
Signal Range: 1,2 0 1 2 3 4 5 6<br />
Full-Scale Current: 4 2.51 nA 25.1 nA 251 nA 2.51 µA 25.1 µA 251 µA 2.50 mA<br />
A/D Resolution: 126 fA 1.26 pA 12.6 pA 126 pA 1.26 nA 12.6 nA 126 nA<br />
(20,000 Count Precision)<br />
Display Noise Floor: ≤ 8 LSD ≤ 1 LSD ≤ 1 LSD ≤ 1 LSD ≤ 1 LSD ≤ 1 LSD ≤ 1 LSD<br />
(Input Open, Filter Off)<br />
Full-Scale Accuracy: 3 ± 0.1% 0.05% 0.05% 0.05% 0.05% 0.05% 0.05%<br />
(Typical)<br />
Full-Scale Accuracy: 3 ±0.48% .30% .30% .30% .30% .30% .30%<br />
(Worst-Case)<br />
1<br />
Listed signal ranges specify meter capability. Available signal ranges are detector dependent.<br />
2<br />
Maximum measurable signal is detector dependent. See description of detector saturation message “SA”, page 80.<br />
3<br />
After 60 min warm-up, followed by execution of AUTOCAL command. See Section 2.4.4.<br />
4<br />
Full scale current may vary due to AUTOCAL compensation of DC offsets.<br />
Analog Output<br />
Full Scale Voltage: 0 - 2.5V into 50Ω<br />
Accuracy: ± 2.5%<br />
xii