LLO?<br />
Local Lockout Enable Query<br />
Syntax: LLO?<br />
Parameters: None<br />
Function:<br />
The query returns the state of the local lockout.<br />
Returns: <br />
is of type and represents the integer 0, if local lockout is<br />
disabled, or 1, if local lockout is enabled.<br />
Related Commands: LLO<br />
MODE_n<br />
Acquisition Mode Select<br />
Syntax: MODE_n <br />
where n is either A or B<br />
Parameters:<br />
The parameter is of type . Its range depends on the detector<br />
attached to the meter. The valid values for this parameter are listed below.<br />
“DCSNGL” specifies DC single mode<br />
“DCCONT” specifies DC continuous mode<br />
Function:<br />
This command sets the mode to be used to acquiring subsequent readings.<br />
Note that the current units will be changed automatically if they are not<br />
compatible with the new mode.<br />
Related Commands: MODE_n?, UNITS_n, UNITS_n?<br />
MODE_n?<br />
Acquisition Mode Query<br />
Syntax: MODE_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns the acquisition mode currently being used by the meter.<br />
Returns: <br />
is of type in the double quote format.<br />
Related Commands: MODE_n, UNITS_n, UNITS_n?<br />