Table 13 – TRIGGER OUT Operations.<br />
Menu Function Keypad Commands Associated Remote Commands<br />
TRIGGER OUT Edit TRIG ON CMPLT TRIGOUT, TRIGOUT?<br />
TRIG AT FREQ<br />
TRIG ON LEVEL<br />
TRIG OUT OFF<br />
TRIG POLARITY Edit ACTIVE HI, LO TRIGOUTPOL, TRIGOUTPOL?<br />
TRIGGER FREQ Edit FREQ= ddd.ddd Hz TRIGOUTFREQ, TRIGOUTFREQ?<br />
TRIGGER LEVEL Edit LVL= d.dddE±dd W TRIGOUTLVL, TRIGOUTLVL?<br />
2.4.9 Trigger Input<br />
The Model 2832-C’s rear panel TTL external trigger input can be enabled or<br />
disabled, Section 2.3.17, and have its edge polarity set. The edge polarity is<br />
accessed via the EXT TRIGGER IN menu function and can be set to rising or<br />
falling edge triggering.<br />
The external trigger affects both channels and, like the R/S key, acts like an<br />
acquisition trigger when in SNGL modes and acts like a toggled acquisition gate<br />
when in CONT modes.<br />
The external trigger pulse width must be ≥200 ns. In DC SNGL mode, the<br />
delay from the trigger input going active to the start of the A/D conversion is<br />