2.4.5 Attenuator<br />
The Attenuator selects the responsivity value, R λ<br />
, to be the value for the<br />
detector alone or the value for the detector-with-attenuator. When ATTN is<br />
ON the ATTN annunciator is lit and the responsivity of the detector-withattenuator<br />
is used. When ATTN is OFF, the annunciator is off and the detector-alone<br />
responsivity is used. If the detector does not have an attentuator, or<br />
if USR CAL is on (Section 2.4.6) the ATTN setting has no effect.<br />
2.4.6 User Calibration<br />
The Model 2832-C allows one to create a detector responsivity which overrides<br />
the responsivities obtained from the detector’s calibration module.<br />
This allows one to account for the effects of additional optics and filters in the<br />
measurement path. When USR CAL is on, the USR CAL annunciator is lit and<br />
the meter adopts the responsivity value displayed by the editable USR RESP<br />
value. Use the PRESENT RESP function to display the current calibration<br />
module responsivity.<br />
The USR RESP units are the same as the PRESENT RESP units. Table 10 lists<br />
the possible user calibration operations.<br />
Table 10 – User Calibration Operations<br />
Menu Operation Keypad Commands Associated Remote Commands<br />
USR CALIB ON, OFF Edit USR CALIB USRCAL_n, USRCAL_n?<br />
ON, OFF<br />
USR RESP<br />
Edit d.dddE±dd A/W* USRRESP_n, USERRESP_n?<br />
PRESENT RESP ENTER to view RESP_n?<br />
d.dddE±dd A/W*<br />
*These units are not editable.<br />
NOTE<br />
n in a remote command stands for A (channel A) or B (channel B).<br />
2.4.7 DC Sampling<br />
The Model 2832-C incorporates two analog-to-digital, (A/D) converters, one<br />
with 20,000 count resolution and a second with 4096 count resolution. The<br />
user may select which A/D will be used for both channels during DC CONT<br />
and DC SNGL acquisition modes. All other modes use the 4096 count A/D.<br />
The 20,000 count A/D converter can operate at sample rates up to 25 Hz.<br />
When two detector calibration modules are plugged into the meter, the 4,096<br />
count A/D converter can operate at sample rates up to 500 Hz. When one<br />
detector calibration module is plugged into the meter, the 4,096 count A/D<br />
converter can operate at sample rates up to 1000 Hz.<br />
The SAMPLE PREC menu command, Table 11, selects which analog-to-digital<br />
converter is used for both channels. The sampling frequency for both<br />
channels can be adjusted within the limits imposed by the SAMPLE PREC<br />
state, see Table 12. When the SAMPLE PREC state changes, the sampling<br />
frequency defaults to 25 Hz if the existing SAMPLE FREQ is incompatible with<br />
the new SAMPLE PREC state. The SAMPLE FREQ and the SAMPLE PREC<br />
settings affect both channels, not just the display channel.<br />
18