LAMBDA_n<br />
Lambda Select<br />
Syntax: LAMBDA_n <br />
where n is either A or B<br />
Parameters:<br />
The parameter is of type which rounds to an integer<br />
with a range that depends on the detector being used. The units of <br />
are nanometers and should correspond to the wavelength of the light<br />
source being measured.<br />
Function:<br />
This command is used to specify the wavelength of light being measured.<br />
This will insure that the proper responsivity calibration point will be used<br />
when the 2832-C calculates measurement values.<br />
Related Commands: ATTN_n, ATTN_n?, LAMBDA_n?, RESP_n?, USRCAL_n,<br />
USRCAL_n?, USRRESP_n, USRRESP_n?<br />
LAMBDA_n?<br />
Lambda Query<br />
Syntax: LAMBDA_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns the wavelength that corresponds to the responsivity<br />
calibration point currently being used by the 2832-C in measurement calculation.<br />
Returns: <br />
is of type and represents an integer. The units of<br />
are nanometers.<br />
Related Commands: ATTN_n, ATTN_n?, LAMBDA_n, RESP_n?, USRCAL_n,<br />
USRCAL_n?, USRRESP_n, USRRESP_n?<br />
LLO<br />
Local Lockout Enable<br />
Syntax: LLO <br />
Parameters:<br />
The parameter is either 0 or 1. If is 0 then local lock-out is<br />
disabled. If is 1 then local lock-out is enabled.<br />
Function:<br />
This command enables or disables the local lockout.<br />
Returns: None<br />
Related Commands: LLO?<br />
61