(8.5x11) 6000 Manual Cover - Newport Corporation
(8.5x11) 6000 Manual Cover - Newport Corporation
(8.5x11) 6000 Manual Cover - Newport Corporation
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
ATTNSN_n?<br />
Attenuator Serial Number Query<br />
Syntax: ATTNSN_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns the serial number of the attenuator that was calibrated<br />
with the attached detector.<br />
When using an attenuator this serial number must match the serial number of<br />
the attenuator being used or data calculations will be inaccurate.<br />
Returns: <br />
is of type in the double quote format.<br />
If no attenuator serial number is available, the query returns an empty double<br />
quoted string, “”.<br />
Related Commands: CALDATE_n?, DETMODEL_n?, DETSN_n?<br />
AUTO_n<br />
Auto Ranging Enable<br />
Syntax: AUTO_n <br />
where n is either A or B<br />
Parameters:<br />
The parameter is either 0 or 1. If is 0 then manual ranging is<br />
enabled. If is 1 then auto ranging is enabled.<br />
Function:<br />
This command enables or disables auto ranging. If enabled and in a continuous<br />
acquisition mode, a signal range is automatically selected to suit the input<br />
signal. If disabled, the meter needs to be manually ranged.<br />
Related Commands: AUTO_n?, RANGE_n, RANGE_n?<br />
AUTO_n?<br />
Auto Ranging Enable Query<br />
Syntax: AUTO_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns a value showing whether the auto ranging feature is<br />
enabled or disabled.<br />
Returns: <br />
is of type which represents the integer 0 if auto ranging is<br />
disabled or 1 if auto ranging is enabled. If enabled, the meter will auto range<br />
when in continuous acquisition modes.<br />
Related Commands: AUTO_n, RANGE_n, RANGE_n?<br />
49