TRIGOUTLVL<br />
Trigger Out Level Select<br />
Syntax: TRIGOUTLVL <br />
Parameters:<br />
The parameter is of type . The units of this parameter are<br />
Watts.<br />
Function:<br />
This command is used to set the level used by the trigger out level comparator<br />
mode. When the power level is at or above this level the trigger out will<br />
go to its active state.<br />
Related Commands: TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,<br />
TRIGOUTLVL?, TRIGOUTPOL, TRIGOUTPOL?<br />
TRIGOUTLVL?<br />
Trigger Out Level Query<br />
Syntax: TRIGOUTLVL?<br />
Parameters: None<br />
Function:<br />
This query returns the trigger out level used by the trigger out level<br />
comparator mode.<br />
Returns: <br />
is of type in exponent notation with units of Watts.<br />
Related Commands: TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,<br />
TRIGOUTLVL, TRIGOUTPOL, TRIGOUTPOL?<br />
TRIGOUTPOL<br />
Trigger Out Polarity Select<br />
Syntax: TRIGOUTPOL <br />
Parameters:<br />
The parameter is a which can be 0 or 1. If the number<br />
rounds to 0, then the polarity will be active low. If the number is 1, then the<br />
polarity will be active high.<br />
Function:<br />
This command is used to set the polarity of the trigger out output to either<br />
active low or active high.<br />
Related Commands: TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,<br />
TRIGOUTLVL, TRIGOUTLVL?, TRIGOUTPOL?<br />
74