*WAI<br />

Wait to continue<br />

Syntax: *WAI<br />

Parameters: None<br />

Function:<br />

The *WAI command causes the device to wait until all pending operations are<br />

complete before processing any commands waiting in the input queue. Care<br />

should be taken when using this command since it is possible to overflow the<br />

input queue by continuing to write commands to the 2832-C while the *WAI<br />

command is in effect.<br />

Note that the *WAI command will essentially “lock out” the remote interfaces<br />

until all pending operations are complete. The *OPC command and *OPC?<br />

query differ from *WAI in this area, as they allow other commands to be<br />

processed while they are waiting for pending operations to complete.<br />

Once the *WAI command is received, it will be in effect while any of the<br />

following operations are in progress:<br />

Single acquisition in progress<br />

Data store in progress with fixed data store buffer<br />

The meter is ranging<br />

The *WAI command can only be canceled by a GPIB interface device clear or<br />

by turning the meter off and on.<br />

Related Commands: *OPC, *OPC?<br />