REFSEL_n?<br />
Reference Source Query<br />
Syntax: REFSEL_n?<br />
Parameters: None<br />
Function:<br />
This query returns the current reference source that will be used in linear and<br />
logarithmic (dB) relative measurements. Returns:<br />
is of type in the double quote format. The possible values<br />
of are:<br />
“STOREF” - Last stored reference measurement<br />
“OTHERCH” - The other channel<br />
“USRREF” - User defined reference value<br />
Related Commands: REFSEL_n, STOREF_n, USRREF_n, USRREF_n?<br />
RESP_n?<br />
Responsivity Query<br />
Syntax: RESP_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns the responsivity value currently being used by the 2832-C<br />
in measurement calculation.<br />
Returns: <br />
is of type and represents a floating point number. The units<br />
of the responsivity depends upon the type of detector family of the<br />
detector in use. See Table 22 for a listing of responsivity units by detector<br />
family.<br />
Related Commands: ATTN_n, ATTN_n?, LAMBDA_n, LAMBDA_n?,<br />
USRCAL_n, USRCAL_n?, USRRESP_n, USRRESP_n?<br />
RUN<br />
Start Both Channel Acquisition<br />
Syntax: RUN<br />
Parameters: None<br />
Function:<br />
This command initiates data acquisition on both channels using the currently<br />
defined acquisition mode. In single type acquisition modes this command will<br />
cause one reading to be acquired. In continuous type acquisition modes this<br />
command will initiate the continuous acquisition of readings. The STOP<br />
command will terminate acquisition.<br />
Related Commands: STOP, STOP_n, RUN_n<br />
65