SFREQ?<br />

DC Sample Frequency Query<br />

Syntax: SFREQ?<br />

Parameters: None<br />

Function:<br />

This query returns the sample rate used for continuous DC acquisition on<br />

both channels<br />

Returns: <br />

is of type in exponent notation with units of hertz or<br />

samples per second.<br />

Related Commands: SFREQ, SPREC, SPREC?<br />

SPREC<br />

DC Sample Precision Select<br />

Syntax: SPREC <br />

Parameters:<br />

The parameter is of type with acceptable values of 20000 or<br />

4096.<br />

Function:<br />

The 2832-C has a 20,000 count high precision A/D and a 4,096 count A/D. This<br />

command determines which of these A/Ds will be used from both channels to<br />

acquire samples during DC single and DC continuous acquisition modes. This<br />

command affects the available range of DC sample frequencies.<br />

Note that the 4,096 count A/D is used during all other acquisition modes.<br />

Related Commands: SFREQ, SFREQ?, SPREC?<br />

SPREC?<br />

DC Sample Precision Query<br />

Syntax: SPREC?<br />

Parameters: None<br />

Function:<br />

This query returns a string that indicates whether the 20,000 count A/D or the<br />

4,096 count A/D will be used during DC CONT or DC SNGL acquisition modes<br />

on both channels.<br />

Returns: <br />

is of type . It will be 20000 if the 20,000 count A/D is selected<br />

and 4096 if the 4,096 count A/D is selected.<br />

Related Commands: SFREQ, SFREQ?, SPREC?<br />