ZERO_n<br />

Signal Zeroing Enable<br />

Syntax: ZERO_n <br />

where n is either A or B<br />

Parameters:<br />

The parameter is a which can be 0 or 1. If the number is 0,<br />

then readings are not adjusted by the stored zero reference. If the number is<br />

1, then readings are adjusted by the stored zero reference.<br />

Function:<br />

This command enables or disables the zeroing feature. Zeroing causes the<br />

stored zero reference (see STOZERO_n command) to be subtracted from<br />

incoming signal readings before the incoming signal is used in measurement<br />

calculations.<br />

Related Commands: STOZERO_n, ZERO_n?, ZEROVAL_n<br />

ZERO_n?<br />

Signal Zeroing Enable Query<br />

Syntax: ZERO_n?<br />

where n is either A or B<br />

Parameters: None<br />

Function:<br />

This query returns a value showing whether or not zeroing is enabled or<br />

disabled.<br />

Returns: <br />

is of type and represents the integer 0, if zeroing is enabled,<br />

or 1, if zeroing is disabled.<br />

Related Commands: STOZERO_n, ZERO_n, ZEROVAL_n<br />

ZEROVAL_n?<br />

Zero Signal Query<br />

Syntax: ZEROVAL_n?<br />

where n is either A or B<br />

Parameters: None<br />

Function:<br />

This query returns the current value being used for zeroing.<br />

Returns: <br />

is of type in exponential notation. The units of this value<br />

are the numerator of the detector’s responsivity, i.e. amps.<br />

Related Commands: STOZERO_n, ZERO_n, ZERO_n?<br />

78