ZERO_n<br />
Signal Zeroing Enable<br />
Syntax: ZERO_n <br />
where n is either A or B<br />
Parameters:<br />
The parameter is a which can be 0 or 1. If the number is 0,<br />
then readings are not adjusted by the stored zero reference. If the number is<br />
1, then readings are adjusted by the stored zero reference.<br />
Function:<br />
This command enables or disables the zeroing feature. Zeroing causes the<br />
stored zero reference (see STOZERO_n command) to be subtracted from<br />
incoming signal readings before the incoming signal is used in measurement<br />
calculations.<br />
Related Commands: STOZERO_n, ZERO_n?, ZEROVAL_n<br />
ZERO_n?<br />
Signal Zeroing Enable Query<br />
Syntax: ZERO_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns a value showing whether or not zeroing is enabled or<br />
disabled.<br />
Returns: <br />
is of type and represents the integer 0, if zeroing is enabled,<br />
or 1, if zeroing is disabled.<br />
Related Commands: STOZERO_n, ZERO_n, ZEROVAL_n<br />
ZEROVAL_n?<br />
Zero Signal Query<br />
Syntax: ZEROVAL_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns the current value being used for zeroing.<br />
Returns: <br />
is of type in exponential notation. The units of this value<br />
are the numerator of the detector’s responsivity, i.e. amps.<br />
Related Commands: STOZERO_n, ZERO_n, ZERO_n?<br />
78