Responsivity Map(s)<br />

This process scales the input value in accordance with current calibration<br />

wavelength and the responsivity map downloaded from the detector calibration<br />

module. The output of this process, i.e. the measurement value, is the<br />

digitized input value divided by the responsivity associated with the current<br />

calibration wavelength or the user defined calibration value. Different<br />

responsivities are used depending upon if the attenuator, ATTN, annunciator is<br />

lit, Section 2.4.5.<br />

Units Correction<br />

Units correction adjusts a measurement value to account for the display units<br />

selected. When the display units are equal to the detector signal units, i.e.<br />

equal to the numerator of the responsivity units, the measurement value is not<br />

adjusted. Otherwise the digitized value is adjusted to account for detector<br />

responsivity and/or additional unit conversions such as W to W/cm 2 .<br />

NOTE<br />

Per area unit conversions such as W to W/cm 2 divide the measurement value<br />

by the active area of the detector. This calculation assumes that the entire<br />

active area of the detector is uniformly illuminated. Per area measurements<br />

where the entire detector active area is not uniformly lit are not accurate. The<br />

user must insure that these conditions are met before utilizing per area units.<br />

3.4 Typical Detector Signals<br />

The flexibility of the Model 2832-C analog signal flow is required in order that<br />

it may properly measure the signals that various types of detectors make.<br />

Basic optical power measurements are related to a measured detector signal,<br />

S, in the following way:<br />

P<br />

S<br />

=<br />

Rλ<br />

Where: R λ<br />

= Detector responsivity at λ .<br />

S = Detector signal<br />

<strong>Newport</strong> Low-Power semiconductor detectors provide a current signal. The<br />

2832-C is capable of 100fA resolution in order to provide the highest sensitivity<br />

performance with these detectors.<br />

3.5 Analog Output<br />

The Model 2832-C provides a 0 to 2.5 volt, into 50Ω,BNC analog output for<br />

signal monitoring. The analog output is the actual amplified detector signal<br />

and is uncorrected for the effects of the detector’s responsivity, AUTO CAL,<br />

and ZERO.<br />

27