*RCL<br />

Recall Meter Configuration<br />

Syntax: *RCL <br />

Parameters:<br />

is of type which rounds to an integer in the range 0 to 9.<br />

This number specifies a configuration storage buffer. The number 0 stands<br />

for the default configuration buffer.<br />

Function:<br />

This command configures the meter to a configuration previously stored in<br />

non-volatile memory (see *SAV), or to the default configuration for the<br />

attached detector. This command will fail if the requested configuration is<br />

incompatible with the detector attached to the meter or empty. The parameters<br />

affected by this command are those listed in Table 8.<br />

Returns: None<br />

Related Commands: *RST, *SAV<br />

*RST<br />

Reset<br />

Syntax: *RST<br />

Parameters: None<br />

Function:<br />

This sets the meter to the default configuration, see Table 8, for the detector<br />

attached to it. It also cancels any outstanding *OPC or *OPC? commands.<br />

Related Commands: *RCL, *SAV<br />

*SAV<br />

Save Meter Configuration<br />

Syntax: *SAV <br />

Parameters:<br />

is of type which rounds to an integer in the range 1 to 9.<br />

This number specifies a configuration storage buffer.<br />

Function:<br />

This command saves the configuration of the meter to the configuration<br />

specified by the parameter. The parameters saved by this command are<br />

those listed in Table 8.<br />

Related Commands: *RCL, *RST<br />