*RCL<br />
Recall Meter Configuration<br />
Syntax: *RCL <br />
Parameters:<br />
is of type which rounds to an integer in the range 0 to 9.<br />
This number specifies a configuration storage buffer. The number 0 stands<br />
for the default configuration buffer.<br />
Function:<br />
This command configures the meter to a configuration previously stored in<br />
non-volatile memory (see *SAV), or to the default configuration for the<br />
attached detector. This command will fail if the requested configuration is<br />
incompatible with the detector attached to the meter or empty. The parameters<br />
affected by this command are those listed in Table 8.<br />
Returns: None<br />
Related Commands: *RST, *SAV<br />
*RST<br />
Reset<br />
Syntax: *RST<br />
Parameters: None<br />
Function:<br />
This sets the meter to the default configuration, see Table 8, for the detector<br />
attached to it. It also cancels any outstanding *OPC or *OPC? commands.<br />
Related Commands: *RCL, *SAV<br />
*SAV<br />
Save Meter Configuration<br />
Syntax: *SAV <br />
Parameters:<br />
is of type which rounds to an integer in the range 1 to 9.<br />
This number specifies a configuration storage buffer.<br />
Function:<br />
This command saves the configuration of the meter to the configuration<br />
specified by the parameter. The parameters saved by this command are<br />
those listed in Table 8.<br />
Related Commands: *RCL, *RST<br />
44