Section 6<br />

Maintenance, Test and Troubleshooting<br />

6.1 Maintenance Procedures<br />

In cleaning the body of this instrument, use only a mild soap and water<br />

solution on a damp cloth.<br />

CAUTION<br />

Do not use acetone or other organic solvents (other than alcohol) on the<br />

Model 2832-C Dual-Channel Power Meter. Organic solvents attack the<br />

paint.<br />

6.2 Power Up Self Test<br />

The Model 2832-C executes a system self test after every power up or reset<br />

and indicates on the display whether the self test has been passed. If the self<br />

test is not passed the display will indicate one of the following error messages:<br />

“CAL MODULE A” “MISSING OR BAD”<br />

“CAL MODULE B” “MISSING OR BAD”<br />

“CAL MODULE A” “CHECKSUM ERROR”<br />

“CAL MODULE B” “CHECKSUM ERROR”<br />

“REGISTER ERR A”<br />

“REGISTER ERR B”<br />

“INTG A/D ERR A”<br />

“INTG A/D ERR B”<br />

“SA A/D ERR A”<br />

“SA A/D ERR B”<br />

If the calibration module messages appear, check to see that the module is<br />

properly inserted.<br />

Should any other message appear, call <strong>Newport</strong> <strong>Corporation</strong>’s Customer<br />

Service Department.<br />