The display shows:<br />
“CAL MODULE A”<br />
Make sure the CAL MODULE is seated<br />
or “CAL MODULE B”<br />
correctly and reset the meter. Try another<br />
and “MISSING OR BAD”<br />
CAL MODULE, if one is available, to<br />
localize the problem to the first<br />
CAL MODULE.<br />
If the first CAL MODULE does not work<br />
after proper seating and reset, but the<br />
second CAL MODULE works, call <strong>Newport</strong><br />
Customer Service and arrange for a repair<br />
of the first CAL MODULE.<br />
If neither the first or second CAL MOD-<br />
ULE works, call <strong>Newport</strong> Customer<br />
Service and arrange for a meter repair.<br />
Be sure to also return the first CAL<br />
MODULE if you cannot otherwise determine<br />
that it is working.<br />
The display shows:<br />
“CAL MODULE A”<br />
Call <strong>Newport</strong> Customer Service and or<br />
“CAL MODULE B”<br />
arrange to return the meter for repair.<br />
and “CHECKSUM ERROR”<br />
The display shows:<br />
“BATTERY FAILURE”<br />
Call <strong>Newport</strong> Customer Service and<br />
arrange to return the meter for battery<br />
replacement.<br />
RS-232 communication does Check the RS-232 cable connection and<br />
not seem to work<br />
cable pinouts, Figure 10. Make sure that<br />
the device talking to the meter is setup<br />
for 8 data bits, no parity, and 1 stop bit.<br />
Also check the echo mode state.<br />
No listener error when attempting Check the GPIB cable connections and<br />
GPIB communications<br />
the GPIB address.<br />
Display value does not change. Press the R/S, R/S A, or R/S B key.<br />
Reading is different than<br />
See Table 19, Common Measurement<br />
expected. Errors. (Section 3.10.6.)<br />
81