5.3 Device Dependent Commands<br />

ATTN_n<br />

Attenuator Calibration Data Enable<br />

Syntax: ATTN_n <br />

where n is either A or B<br />

Parameters:<br />

The parameter is either 0 or 1. If is 0, then detector<br />

responsivity data is used to calculate readings. If is 1, then the<br />

detector + attenuator responsivity data is used to calculate readings.<br />

Function:<br />

This command enables or disables the use of attenuator responsivity data<br />

when user calibration is disabled. If enabled, responsivity data collected with<br />

the attenuator attached to the detector is used in data calculations. If disabled,<br />

responsivity data collected with the detector alone is used in data<br />

calculations.<br />

This command will generate an error if the attached detector does not<br />

support an attenuator.<br />

Related Commands: ATTN_n?, ATTNSN_n?, LAMBDA_n,<br />

LAMBDA_n?, RESP_n?, USRCAL_n, USRCAL_n?, USRRESP_n,<br />

USRRESP_n?<br />

ATTN_n?<br />

Attenuator Calibration Data Enable Query<br />

Syntax: ATTN_n?<br />

where n is either A or B<br />

Parameters: None<br />

Function:<br />

This query returns a value showing whether or not attenuator responsivity<br />

data will be used when user calibration is disabled.<br />

Returns: <br />

is of type which represents the integer 0 if detector<br />

responsivity data is being used or 1 if detector + attenuator responsivity data<br />

is being used.<br />

Related Commands: ATTN_n, ATTNSN_n?, LAMBDA_n, LAMBDA_n?,<br />

RESP_n?, USRCAL_n, USRCAL_n?, USRRESP_n, USRRESP_n?<br />

48