5.3 Device Dependent Commands<br />
ATTN_n<br />
Attenuator Calibration Data Enable<br />
Syntax: ATTN_n <br />
where n is either A or B<br />
Parameters:<br />
The parameter is either 0 or 1. If is 0, then detector<br />
responsivity data is used to calculate readings. If is 1, then the<br />
detector + attenuator responsivity data is used to calculate readings.<br />
Function:<br />
This command enables or disables the use of attenuator responsivity data<br />
when user calibration is disabled. If enabled, responsivity data collected with<br />
the attenuator attached to the detector is used in data calculations. If disabled,<br />
responsivity data collected with the detector alone is used in data<br />
calculations.<br />
This command will generate an error if the attached detector does not<br />
support an attenuator.<br />
Related Commands: ATTN_n?, ATTNSN_n?, LAMBDA_n,<br />
LAMBDA_n?, RESP_n?, USRCAL_n, USRCAL_n?, USRRESP_n,<br />
USRRESP_n?<br />
ATTN_n?<br />
Attenuator Calibration Data Enable Query<br />
Syntax: ATTN_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns a value showing whether or not attenuator responsivity<br />
data will be used when user calibration is disabled.<br />
Returns: <br />
is of type which represents the integer 0 if detector<br />
responsivity data is being used or 1 if detector + attenuator responsivity data<br />
is being used.<br />
Related Commands: ATTN_n, ATTNSN_n?, LAMBDA_n, LAMBDA_n?,<br />
RESP_n?, USRCAL_n, USRCAL_n?, USRRESP_n, USRRESP_n?<br />