RANGE_n<br />

Signal Range Select<br />

Syntax: RANGE_n <br />

where n is either A or B<br />

Parameters:<br />

The parameter is of type . The range of this parameter will<br />

depend on the detector and mode being used to acquire data. See the section<br />

entitled “Electrical Specifications” at the beginning of this manual for the<br />

signal range that corresponds to the current setting of .<br />

Function:<br />

This command is used to select a new manual signal range. This command<br />

will disable the auto ranging feature.<br />

Related Commands: AUTO_n, AUTO_n?, RANGE_n?<br />

RANGE_n?<br />

Signal Range Query<br />

Syntax: RANGE_n?<br />

where n is either A or B<br />

Parameters: None<br />

Function:<br />

This query returns a value that represents the current signal range setting<br />

being used whether the meter is auto ranging or in manual ranging mode. See<br />

Section 2.3.11.<br />

Returns: <br />

is of type and represents the gain level currently being<br />

used for either auto or manual ranging. See the section entitled “Electrical<br />

Specifications” at the beginning of this manual for the signal range that<br />

corresponds to .<br />

Related Commands: AUTO_n, AUTO_n?, RANGE_n<br />

REFSEL_n<br />

Reference Source Select<br />

Syntax: REFSEL_n <br />

Parameters:<br />

is of type in the double quote format. The possible values<br />

of are: “STOREF” - Last stored reference measurement, “OTHERCH” -<br />

The other channel, “USRREF” - User defined reference value<br />

Function:<br />

This command is used to select a reference source to be used in linear and<br />

logarithmic (dB) relative measurements. This command will cause an error if<br />

an attempt is made to select the other channel as a reference if it is not<br />

available.<br />

Returns: None<br />

Related Commands: REFSEL_n?, STOREF_n, USRREF_n, USRREF_n?<br />