BARGRAPH<br />
Bar Graph Enable<br />
Syntax: BARGRAPH <br />
Parameters:<br />
The parameter is either 0 or 1. If is 0 then the bargraph is<br />
disabled. If is 1 then the bargraph is enabled.<br />
Function:<br />
This command enables or disables the bargraph feature. If enabled, the<br />
bargraph will be displayed when the units of measurement are not ratio type<br />
of units and the meter is not in INTG mode.<br />
Related Commands: BARGRAPH?<br />
BARGRAPH?<br />
Bar Graph Enable Query<br />
Syntax: BARGRAPH?<br />
Parameters: None<br />
Function:<br />
This query returns a value showing whether the bargraph is enabled or<br />
disabled. Note: The bargraph is not displayed while in either GPIB or RS-232<br />
mode regardless of enabled status.<br />
Returns: <br />
is of type which represents the integer 0 if the bar graph is<br />
disabled and 1 if the bar graph is enabled.<br />
Related Commands: BARGRAPH<br />
CALDATE_n?<br />
Calibration Date Query<br />
Syntax: CALDATE_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns the calibration date of the detector as read from the<br />
memory module attached to the 2832-C.<br />
Returns: <br />
is of type in the double quote format.<br />
Related Commands: ATTNSN_n?, DETMODEL_n?, DETSN_n?<br />