CH?<br />
Available Channel Query<br />
Syntax: CH?<br />
Parameters: None<br />
Function:<br />
This query is used to see which channels are available in the meter. A<br />
channel is available if a calibration module was plugged into that channel on<br />
powerup and passed the self test.<br />
Returns: <br />
is of type in the double quote format. The possible<br />
values of are:<br />
“A” - only channel A is available<br />
“B” - only channel B is available<br />
“AB” - Both channel A and B are available<br />
Related Commands: None<br />
DETMODEL_n?<br />
Detector Model Query<br />
Syntax: DETMODEL_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns the model number of the detector that belongs to the<br />
memory module attached to the specified channel.<br />
Returns: <br />
is of type in the double quote format.<br />
Related Commands: ATTNSN_n?, CALDATE_n?, DETSN_n?<br />
DETSN_n?<br />
Detector Serial Number Query<br />
Syntax: DETSN_n?<br />
where n is either A or B<br />
Parameter: None<br />
Function:<br />
This query returns the serial number of the detector that belongs to the<br />
memory module attached to the specified channel.<br />
This serial number must match the serial number of the detector attached to<br />
the 2832-C.<br />
Returns: <br />
is of type in the double quote format.<br />
Related Commands: ATTNSN_n?, CALDATE_n?, DETMODEL_n?<br />
51