CH?<br />

Available Channel Query<br />

Syntax: CH?<br />

Parameters: None<br />

Function:<br />

This query is used to see which channels are available in the meter. A<br />

channel is available if a calibration module was plugged into that channel on<br />

powerup and passed the self test.<br />

Returns: <br />

is of type in the double quote format. The possible<br />

values of are:<br />

“A” - only channel A is available<br />

“B” - only channel B is available<br />

“AB” - Both channel A and B are available<br />

Related Commands: None<br />

DETMODEL_n?<br />

Detector Model Query<br />

Syntax: DETMODEL_n?<br />

where n is either A or B<br />

Parameters: None<br />

Function:<br />

This query returns the model number of the detector that belongs to the<br />

memory module attached to the specified channel.<br />

Returns: <br />

is of type in the double quote format.<br />

Related Commands: ATTNSN_n?, CALDATE_n?, DETSN_n?<br />

DETSN_n?<br />

Detector Serial Number Query<br />

Syntax: DETSN_n?<br />

where n is either A or B<br />

Parameter: None<br />

Function:<br />

This query returns the serial number of the detector that belongs to the<br />

memory module attached to the specified channel.<br />

This serial number must match the serial number of the detector attached to<br />

the 2832-C.<br />

Returns: <br />

is of type in the double quote format.<br />

Related Commands: ATTNSN_n?, CALDATE_n?, DETMODEL_n?<br />