STOREF_n?<br />
Store Reference Reading Query<br />
Syntax: STOREF_n?<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This query returns the latest reading stored as a reference reading. The units<br />
of the reference reading defaults to Watts.<br />
Returns: <br />
is a floating point number in the exponential format.<br />
Related Commands: REFSEL_n, REFSEL_n?, USRREF_n, USRREF_n?, STOREF_n<br />
STOZERO_n<br />
Store Zero Signal<br />
Syntax: STOZERO_n<br />
where n is either A or B<br />
Parameters: None<br />
Function:<br />
This command takes the latest reading and stores it as a zero signal reference<br />
value to be used when zeroing is enabled. The units of the zero reference<br />
defaults to Amperes.<br />
Related Commands: ZERO_n, ZERO_n?, ZEROVAL_n?<br />
TRIGOUT<br />
Trigger Out Mode Select<br />
Syntax: TRIGOUT <br />
Parameters:<br />
The parameter is of type . The valid values for this parameter<br />
are listed below.<br />
“CMPLT” specifies conversion complete mode<br />
“FREQ” specifies frequency mode<br />
“LVL” specifies level comparator mode<br />
“OFF” specifies off mode<br />
Function:<br />
This command determines in which mode the trigger out output on the back<br />
panel will operate.<br />
Certain modes require additional parameters to be set.<br />
Related Commands: TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,<br />
TRIGOUTLVL, TRIGOUTLVL?, TRIGOUTPOL, TRIGOUTPOL?<br />
72