Q36: To eliminate the decrease in sensitivity close to a wall which is parallel<br />

to the beam direction, the transducer used should be:<br />

A. As small as possible<br />

B. As low frequency as possible<br />

C. Both A & B<br />

D. Large and with a frequency as high as possible<br />

Q45: The length of the near field <strong>for</strong> a 2.5cm diameter, 5MHz transducer<br />

placed in oil with V=1.4 x 10 5 cm/s is approximately:<br />

A. 0.028 cm<br />

B. 6.25 cm<br />

C. 22.3 cm<br />

D. 55.8 cm