• High-voltage or low-voltage driving current, which is selected <strong>for</strong> the<br />

transducer with a transducer voltage switch.<br />

• Gated alarm units, which enable the use of automatic alarms when flaws<br />

are detected. This is accomplished by setting up controllable time spans<br />

on the display that correspond to specific zones within the test piece.<br />

Signals appearing within the gates may automatically operate visual or<br />

audible alarms. These signals may also be passed on to display devices<br />

or strip-chart recorders or to external control devices. Gated alarm units<br />

usually have three controls: the gate-start or delay control, which adjusts<br />

the location of the leading edge of the gate on the oscilloscope trace; the<br />

gate-length control, which adjusts the length of the gate or the location of<br />

the gate trailing edge; and the alarm-level or sensitivity control, which<br />

establishes the minimum echo height necessary to activate an alarm<br />

circuit. A positive/negative logic switch determines whether the alarm is<br />

triggered above or below the threshold level.