Figure D.1: Broadbeam transmission at 511 keV as a function <strong>of</strong> lead thickness<br />

1.0000<br />

Lead<br />

0.1000<br />

Transmission<br />

0.0100<br />

0.0010<br />

0.0001<br />

0 10 20 30 40 50<br />

Thickness (mm)<br />

(Reproduced from AAPM, 2006)<br />

Figure D.2: Broadbeam transmission at 511 keV as a function <strong>of</strong> concrete thickness<br />

1.0000<br />

Concrete<br />

0.1000<br />

Transmission<br />

0.0100<br />

0.0010<br />

0.0001<br />

0 10 20 30 40 50<br />

Thickness (cm)<br />

(Reproduced from AAPM, 2006)<br />

<strong>The</strong> <strong>Design</strong> <strong>of</strong> <strong>Diagnostic</strong> <strong>Medical</strong> <strong>Facilities</strong> <strong>where</strong> Ionising Radiation is used 99