30.01.2015 Views

The Design of Diagnostic Medical Facilities where ... - ResearchGate

The Design of Diagnostic Medical Facilities where ... - ResearchGate

The Design of Diagnostic Medical Facilities where ... - ResearchGate

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>where</strong> Ḋ (0) is the product <strong>of</strong> the dose rate constant for the patient (0.092 μSvm 2 /MBqh for 18 F) and the<br />

administered activity A 0<br />

(MBq) and R tU<br />

is the reduction factor for the uptake time.<br />

Imaging room<br />

To calculate the dose from a patient during imaging, the decay during the uptake phase must be taken into<br />

account. In addition the reduction factor <strong>of</strong> 0.85 as a result <strong>of</strong> patient voiding should also be taken into<br />

account. <strong>The</strong> total dose (μSv) at a distance d (m) from the patient is given by<br />

F<br />

D(t<br />

i<br />

) = D(0)<br />

. × ti<br />

× Rti<br />

× 0.85 ×<br />

d<br />

u<br />

2<br />

Equation 5.11<br />

<strong>where</strong> Ḋ (0) is the product <strong>of</strong> the dose rate constant for the patient (0.092 μSv m 2 /MBqh) for 18 F and the<br />

administered activity A 0<br />

(MBq), R ti<br />

is the reduction factor for the imaging time t i<br />

and F u<br />

is the decay in the<br />

radionuclide activity since administration.<br />

To calculate the annual dose at a boundary, the calculated dose should be multiplied by the annual workload.<br />

<strong>The</strong> maximum allowable transmission can then be calculated using the dose constraint and occupancy<br />

factors in a similar manner to that employed in X‐ray.<br />

Shielding factors<br />

A variety <strong>of</strong> attenuation coefficients has been used to estimate transmission requirements for PET facilities.<br />

In some cases narrow-beam, good geometry attenuation coefficients for lead and concrete have been used.<br />

However, calculations based on these values will not provide sufficient shielding since they neglect scatter<br />

buildup factors. <strong>The</strong> AAPM recommends using values <strong>of</strong> broad beam transmission factors for lead, concrete,<br />

and iron that are based on consistent Monte Carlo calculations. Plots <strong>of</strong> the broad beam transmission at 511<br />

keV are given in Appendix D for lead and concrete. <strong>The</strong> transmission factors for 18 F through lead and concrete<br />

are also given.<br />

5.3 Examples <strong>of</strong> shielding calculations<br />

Examples <strong>of</strong> shielding calculations for radiology, nuclear medicine and PET facilities are presented in Sections<br />

5.3.1 and 5.3.2. <strong>The</strong> aim <strong>of</strong> the radiology examples is to show how the BIR and NCRP methodologies may be<br />

applied in particular cases using the dose constraints in Table 2.1 and other local data. In practice it may be<br />

prudent in marginal cases to calculate the shielding requirements using both methodologies. <strong>The</strong> examples are<br />

provided for illustrative purposes only and are not intended to substitute for the RPA’s assessment.<br />

<strong>The</strong> <strong>Design</strong> <strong>of</strong> <strong>Diagnostic</strong> <strong>Medical</strong> <strong>Facilities</strong> <strong>where</strong> Ionising Radiation is used 65

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!