30.01.2015 Views

The Design of Diagnostic Medical Facilities where ... - ResearchGate

The Design of Diagnostic Medical Facilities where ... - ResearchGate

The Design of Diagnostic Medical Facilities where ... - ResearchGate

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Example 3: X‐ray room window (BIR method)<br />

<strong>The</strong> general radiographic room is located on the first floor. <strong>The</strong>re is a window in the external wall, which is<br />

at a height much greater than 2 m from the outside ground. <strong>The</strong> distance from the patient table to the<br />

window is 2 m. An occupant works at the window <strong>of</strong> an <strong>of</strong>fice across the courtyard at a distance <strong>of</strong> 8 m from<br />

the window <strong>of</strong> the X-ray room.<br />

<strong>The</strong> weekly workload for the room has been obtained from the Radiology Information System (RIS) and<br />

found to be 1000 Gy cm 2 for 300 examinations, with an average X‐ray beam energy <strong>of</strong> 100 kVp.<br />

What shielding is required for this window<br />

<strong>The</strong>re is no vertical bucky in the room and lateral exposures will not be directed towards the window in this<br />

instance, so only scattered radiation needs to be considered.<br />

<strong>The</strong> annual dose constraint, D c<br />

= 0.3 mSv<br />

Occupancy in <strong>of</strong>fice across courtyard, T = 100%<br />

Distance from patient on table to<br />

occupant in the <strong>of</strong>fice, d<br />

= 10 m<br />

Average weekly DAP = 1000 Gy cm 2<br />

Scatter factor at 1 m, S max<br />

= [(0.031 x kVp) + 2.5] μGy (Gy cm 2 ) -1 (Equation 5.3)<br />

= 5.6 μGy (Gy cm 2 ) -1<br />

<strong>The</strong> incident kerma,<br />

S<br />

max<br />

× DAP<br />

K<br />

inc<br />

=<br />

d<br />

2<br />

(Equation 5.4)<br />

5.6×<br />

1000<br />

= 56μGy = 0.056 mGy ≡ 0.056 mSv<br />

10<br />

K<br />

inc<br />

=<br />

2<br />

Max. transmission,<br />

B<br />

=<br />

K<br />

inc<br />

D<br />

c<br />

× T × 52<br />

= 0.3 mSv<br />

0.056 mSv × 1×<br />

52<br />

= 0.1030<br />

From equation 5.6 for the transmission through lead and concrete, 0.3 mm lead or 26 mm concrete is<br />

needed to shield the window (see also Appendix C) (neglecting any attenuation in the glass<br />

window).<br />

Window shielding is therefore required in this room even though the window is located at a height <strong>of</strong><br />

greater than 2 m from the outside ground. Shielding may be provided as outlined in Section 6.3.4.<br />

Should there be a requirement to direct lateral exposures towards the window then the shielding<br />

requirement will be increased.<br />

If there are two adjacent X‐ray rooms with external windows facing into the courtyard then the<br />

combined effect <strong>of</strong> both rooms on the <strong>of</strong>fice will have to be considered.<br />

<strong>The</strong> <strong>Design</strong> <strong>of</strong> <strong>Diagnostic</strong> <strong>Medical</strong> <strong>Facilities</strong> <strong>where</strong> Ionising Radiation is used 69

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!