(RS-232 Personal <strong>Co</strong>mputer, continued)<br />

If the Accu-Setter is executing the MPF FXT program, then the response packet will just<br />

contain the displayed fixture. The response packet will contain five fields for each feature<br />

being measured. Each field will be separated by a semicolon (;) delimiter. If their is no<br />

data for a particular field, the delimiter will still be included in the response packet. The<br />

following five fields are allocated for each measured feature.<br />

Field 1 — Check Name, 8 ASCII Characters<br />

Field 2 — Check Value, 5 - 8 ASCII Characters (+0.12345)<br />

Field 3 — Scale, 2 ASCII Characters (in. or mm)<br />

Field 4 — Check Status, 0 - 7 ASCII Characters (over, +apprch, good, -apprch, under)<br />

Field 5 — Check Class, 0 - 8 ASCII Characters (class xx)<br />

The response packet can vary in length depending on the number of features programmed<br />

in the Accu-Setter. the end of each response packet will contain the following five bytes:<br />

Example:<br />

1 2 3 4 5<br />

<br />