(RS-232 Personal <strong>Co</strong>mputer, continued)<br />
If the Accu-Setter is executing the MPF FXT program, then the response packet will just<br />
contain the displayed fixture. The response packet will contain five fields for each feature<br />
being measured. Each field will be separated by a semicolon (;) delimiter. If their is no<br />
data for a particular field, the delimiter will still be included in the response packet. The<br />
following five fields are allocated for each measured feature.<br />
Field 1 — Check Name, 8 ASCII Characters<br />
Field 2 — Check Value, 5 - 8 ASCII Characters (+0.12345)<br />
Field 3 — Scale, 2 ASCII Characters (in. or mm)<br />
Field 4 — Check Status, 0 - 7 ASCII Characters (over, +apprch, good, -apprch, under)<br />
Field 5 — Check Class, 0 - 8 ASCII Characters (class xx)<br />
The response packet can vary in length depending on the number of features programmed<br />
in the Accu-Setter. the end of each response packet will contain the following five bytes:<br />
Example:<br />
1 2 3 4 5<br />