4.6 Automatic Air Shutoff (Optional)<br />
The optional automatic air shutoff can be used with any air gaging application. The<br />
automatic air shutoff will shut the air flow to the tooling to a minimal level when the<br />
tooling is not being used to measure a part and return the air flow to the normal operating<br />
level when the tooling is place in gage position in a part.<br />
NOTE: The auto air shutoff only operates in the RUN mode. In calibration mode the<br />
auto air shutoff is disabled so air inputs can be set up.<br />
Required Hardware:<br />
• Auto Air Shutoff Assembly #5913250<br />
Installation<br />
1) Turn off the power to the <strong>Accusetter</strong> II.<br />
2) Remove the existing filter/regulator assembly from the back of the <strong>Accusetter</strong> II and<br />
install the filter/regulator assembly with the shutoff valve, Edmunds #5911200.<br />
3) Plug 25 pin connector on the shutoff cable, Edmunds #5911018, into I/O port "B" on<br />
the rear of the <strong>Accusetter</strong> II and connect the other end to the solenoid valve.<br />
4) Turn on the power to the <strong>Accusetter</strong> II.<br />
5) Proceed to the Setup section below.<br />
4-25