2.7 E9310 (2) Channel LVDT Signal <strong>Co</strong>nditioning Module<br />
The E9310 module is a two-channel signal conditioning amplifier for inductive type<br />
transducers such as LVDTs that converts the outputs of the transducers into a conditioned<br />
signal for the main controller board. The module is mounted in the lower bay of the<br />
<strong>Accusetter</strong> II.<br />
Input magnification must be set using jumpers allowing magnification reduction to be set<br />
to 10x for long range transducers or 1x for standard transducers.<br />
A jumper matrix on the board allows it to accept signals in or send signals out to the<br />
analog output connector. By placing the jumper pin for the desired signal line on one of<br />
the six buss lines, the signal can now be sent or received by any other units connected to<br />
the buss.<br />
Power to the column must be turned off prior to installing or removing a signal<br />
conditioning module.<br />
Refer to the Advanced Operation, LVDT Module Setup section for additional<br />
information on jumper settings.<br />
Figure 2.7 - E9310 (2) Channel LVDT Signal <strong>Co</strong>nditioning Module<br />
2-11