3.2 Signal <strong>Co</strong>nditioning Module Setup<br />

For LVDT modules E9310 & E9320 attenuation jumpers must be set to either 1x for<br />

standard transducers or 10x for long range transducers. Refer to the "Advanced<br />

Operations" section for additional information.<br />

For either an LVDT or A/E signal conditioning module a jumper matrix on the board<br />

allows it to accept signals in or send signals out to the analog output connector. Refer to the<br />

"Advanced Operations" section of the manual for additional information on setting these<br />

jumpers.<br />

Signal <strong>Co</strong>nditioning Module Auto Recognition<br />

When an E9310, E9320, E9330, or E9340 module is installed in the lower bay the<br />

<strong>Accusetter</strong> II will automatically detect the type of signal conditioning module installed and<br />

certain setup parameters will be programmed automatically based on the jumper settings on<br />

the module. This feature is not available on legacy signal conditioning modules.<br />

For an E9310 or E9320 module the following parameters will be detected:<br />

• Under the "Xducer" menu the Signal <strong>Co</strong>nditioning will be set to "Electric"<br />

• The position of the attenuation jumpers J5-J8 and J10 (1x or 10X) will be detected<br />

For an E9330 or E9340 module the following parameters will be detected:<br />

• Under the "Xducer" menu the Signal <strong>Co</strong>nditioning will be set to "Air"<br />

When the <strong>Accusetter</strong> II is powered on the module ID code of the currently installed module<br />

is displayed in the scroll of start up information. For the LVDT modules E9310 & E9320<br />

the ID code will also identify the status of the "J10" attenuation jumper (1x or 10x).<br />

Module<br />

ID <strong>Co</strong>de<br />

E9010, 20, 30, 40 Legacy Module<br />

FF<br />

(no module recognition)<br />

E9310, Jumper J10 = 1x E1<br />

E9310, Jumper J10 = 10x E2<br />

E9320, Jumper J10 = 1x E3<br />

E9320, Jumper J10 = 10x E4<br />

E9330<br />

E5<br />

E9340<br />

E6<br />

Power to the column must be turned off prior to installing or removing a signal<br />

conditioning module.<br />

3-6