Zero Master Calibration<br />
1) From RUN mode, press the CAL button.<br />
2) Rotate the ENTER button until "ZERO" is displayed and then press ENTER. "INS<br />
ZERO" will be displayed.<br />
NOTE: All checks that are programmed with a ZERO master will be turned "ON"<br />
by default and will therefore be gaged at the same time. If the ZERO masters are to<br />
be gaged separately then the checks must be turned "OFF" manually from the "CHK<br />
SEL" menu.<br />
3) Place the ZERO master into gage position and press ENTER. "GAGE ZERO" will be<br />
displayed.<br />
4) Press the ENTER button. The ZERO master will be gaged and then the display will<br />
update to "MASTERED" if the calibration was successful. If the calibration was<br />
successful the gage is ready to measure parts, see Operation. Remove the ZERO<br />
master and press RUN. If the calibration was not completed successfully then see the<br />
Troubleshooting section of this manual.<br />
3-50