PROGRAMMING - SYSTEM OPTIONS<br />
Buss In/Buss Out Setup<br />
Note: For multiple check and multiple fixture programs BUSS IN and BUSS OUT<br />
options are available, for column to column communications. These signal are used<br />
with the I/O "A" and "B" 25 pin connections.<br />
1) From the COMM menu rotate the enter button to display "BUS OUT" and then<br />
press ENTER.<br />
2) Rotate the ENTER button to select the name of the check or fixture to be bussed<br />
out and then press ENTER.<br />
3) "BUS OUT" will display. Rotate the enter button to display "BUS IN" and<br />
press the ENTER button.<br />
4) Rotate the ENTER button to select "INPUT C" or "INPUT D" and press the<br />
ENTER button.<br />
5) Rotate the ENTER button to select "AIR" if the signal to be bussed in is from an<br />
air gage or "ELEC" if the signal to be bussed in is from an electronic gage.<br />
6) Press the ENTER button.<br />
7) Press the RST button. If a second bussed in signal is to be set up repeat steps 4<br />
- 6.<br />
8) When all inputs have been setup press the RUN button.<br />
9) Rotate the ENTER button to display "SAVE" and then press ENTER to save the<br />
current setup.<br />
3-41