Attenuation Jumpers<br />

Jumpers J3, J4, and J10 allow the input attenuation to be set to 10x or 1x. For standard<br />

transducers the jumpers should be set to 1x. If long range transducers are to be used the<br />

jumpers should be set to 10x.<br />

When an E9310 or E9320 module is used on an <strong>Accusetter</strong> II the column will<br />

automatically recognize the setting of the 1X/10X jumper and program the column<br />

accordingly. See figure 4.6.<br />

NOTE: Jumpers J5, J6, J7, J8, and J10 must all be set to the same value.<br />

Figure 4.6 - Attenuation Jumpers<br />